First page Back Continue Last page Graphics
Windowless High Throughput
5ID-D has a large (12” 6-way cross) vacuum chamber equipped with wheel-based sample changer. The chamber provides a windowless path starting downstream of the incident beam intensity monitor, through 3 sets of collimating slits, and ending just before the detector. With this setup we have acheived background noise levels < 10 cps above the detector noise of our MarCCD detector (a 16 bit, 162 mm CCD detector from MarUSA).
52 and 86 hole wheels are available for sample mounting. The sample changer can thus provide a throughput of hundreds of static samples a day.